THERMOCOUPLE ACCURACY SPECIFICATIONS
Type Range ... Accuracy ...
Measure (4 wire) Source
Ni 120 -80 °C to +260 °C 0.2 °C 0.2 °C
Pt 100 - 385 -200 °C to +800 °C 0.33 °C 0.33 °C
Pt 100 - 3926 -200 °C to +630 °C 0.3 °C 0.3 °C
Pt 100 - 3916 -200 °C to +630 °C 0.3 °C 0.3 °C
Pt 200 - 385 -200 °C to +250 °C
250 °C to 630 °C
0.2 °C
0.8 °C
0.2 °C
0.8 °C
Pt 500 - 385 -200 °C to +500 °C
500 °C to +630 °C
0.3 °C
0.4 °C
0.3 °C
0.4 °C
Pt 1000 - 385 -200 °C to +100 °C
100 °C to +630 °C
0.2 °C
0.3 °C
0.2 °C
0.2 °C
Resolution
RTD 0.1 °C, 0.1 °F
UNIVERSAL
MEASURING DEVICE
THERMOCOUPLE ACCURACY
SPECIFICATIONS
Thermocouple Measure
or
source
J -200 to 0 °C
0 to +1200 °C
1.0 °C
0.7 °C
K -200 to 0 °C
0 to +1370 °C
1.2 °C
0.8 °C
T -200 to 0 °C
0 to +400 °C
1.2 °C
0.8 °C
E -200 to 0 °C
0 to +950 °C
0.9 °C
0.7 °C
R
-20 to 0 °C
0 to +500 °C
+500 to
+1750 °C
2.5 °C
1.8 °C
1.4 °C
S
-20 to 0 °C
0 to +500 °C
+500 to
+1750 °C
2.5 °C
1.8 °C
1.5 °C
B
+600 to
+800 °C
+800 to
+1000 °C
+1000 to
+1800 °C
2.2 °C
1.8 °C
1.4 °C
L
-200 to 0 °C
0 to +900 °C
0.85 °C
0.7 °C
U
-200 to 0 °C
0 to +400 °C
1.1 °C
0.75 °C
N
-200 to 0 °C
0 to +1300 °C
1.5 °C
0.9 °C
Resolution
J, K, T, E, L, N, U
B, R, S
0.1 °C.
0.1 °F
1 °C, 1 °F
Notes
Accuracy specications include 0.2 °C cold
junction uncertainty.
FEATURES
Simultaneous Function Capability
Channel A
Channel B
24,000 mA DC M M or S
24,000 mA DC M
100,00 mV DC M or S
30,000 V DC Measure M
20,000 V DC Measure
10,000 V DC Source
M or S
Source: 15 to 3200 Ω, Measure: 0 to 3200 Ω M or S
Thermocouple J, K, E, R, S, B, L, U, N M or S
RTD Ni120; Pt100 (392); Pt100 (JIS);
Pt100, 200, 500, 1000 (385)
M or S
Pressure (using 700PEx modules) M M used as S
Frequency; Squarewave, 1 CPM to 10 kHz;
xed amplitude 5 V p-p
M or S
M = Measure S = Source/Simulate